The system cannot achieve the required negative pressure, or loses vacuum too quickly during a test.

Approximately 46 x 45.5 x 61 cm, weighing around 45–48 kg when empty. autoclave euronda e9 service manual

Debris can prevent EV1 or EV2 from closing completely. Disassemble, clean the plungers, or replace the valve seats. The system cannot achieve the required negative pressure,

Run the pump independently to measure maximum negative pressure capabilities (should drop below -0.80 bar). clean the plungers

Separate clean water and used water internal inspection tanks. Main Hydraulic Components

The physical relationship between saturated steam pressure and temperature is uncoupled (indicates non-condensable gases in the chamber).